Product Overview

AQ6374 Overview and Applications

The Yokogawa AQ6374 measures spectra across visible and near-infrared wavelengths. It is suited to laser and broadband-source characterization and wavelength-dependent fiber-loss measurements. Eight resolution settings let engineers balance spectral detail and measurement needs.

Measurement and Analysis

Seven traces can be displayed together for comparison. Built-in analysis covers laser, LED, optical-filter, and color measurements. A purge connection helps reduce water-vapor absorption within the monochromator. The internal reference source supports optical alignment and wavelength calibration.

Applications and Automation

  • Characterize visible-to-telecom laser sources and broadband emitters.
  • Measure optical-fiber loss versus wavelength with a suitable external source.
  • Evaluate visible LED spectra and color.
  • Integrate remote measurements using GP-IB, RS-232, or Ethernet; Ethernet supports TCP/IP and VXI-11 or socket control.

Model and Configuration Notes

Yokogawa lists the AQ6374 as discontinued in November 2023 and identifies the AQ6374E as its successor. This listing describes the AQ6374. Confirm the unit’s connector adapters, supplied accessories, condition, and calibration records with your quotation.

Specification conditions: [A] 0.05 nm resolution. [A–C] 9.5/125 µm SM fiber after internal alignment, without purge gas. [B] Absolute level display, resolution ≥0.2 nm. [C] Specified PC-polished SM fiber. Sensitivity and sweep figures assume pulse mode off. [D] 1,001 points, one average; excludes 570–580 and 900–1000 nm. Performance temperature: 18–28°C. Follow the English guide for complete conditions, warm-up, and calibration.

Confirm the AQ6374 configuration

Request photos of the optical input and reference-source output, the fitted adapters, and any test or calibration records. Share your wavelength, fiber type, required resolution and calibration needs.

Ask about availability and request unit photos or test records · Buying FAQ

Model FAQ

Which optical adapters should I confirm?

AQ9447 fits the optical input and AQ9441 fits the reference-source output. FC and SC types are available. Confirm the connector and polish for each port.

Can I move my setup to the AQ6374E?

Both models cover 350 to 1750 nm. The AQ6374 supports RS-232C. The AQ6374E documents GP-IB and Ethernet. Check your connectors, fixtures and automation before substituting.

Manufacturer documents

Technical Information

AQ6374 Technical Specifications

Optical Performance

Wavelength range
350–1750 nm
Resolution settings
0.05, 0.1, 0.2, 0.5, 1, 2, 5, and 10 nm
Wavelength accuracy
±0.2 nm overall; ±0.05 nm at 1523 nm and at 633 nm after 633 nm HeNe calibration [A]
Wavelength repeatability
±0.015 nm over 1 minute [A]
Minimum wavelength sampling interval
0.002 nm
Sampling points
101–100,001; AUTO available
Level sensitivity
−70 dBm at 400–900 nm; −80 dBm at 900–1600 nm; HIGH3 [B]
Maximum safe total input
+10 dBm at 400–550 nm; +20 dBm at 550–1750 nm [B]
Level accuracy
±1.0 dB at 1550 nm, −20 dBm input, HIGH1–HIGH3 [B,C]
Level linearity
±0.2 dB for −40 to 0 dBm input, HIGH1–HIGH3 [B]
Close-in dynamic range
60 dB at peak ±1 nm (633/1523 nm); 0.05 nm resolution [A]
Sweep time
0.5 s at NORM_AUTO, span ≤100 nm [D]

Fiber Input and Bench Integration

Supported fibers
SM 9.5/125 µm; GI 50/125 and 62.5/125 µm; large-core fibers up to 800 µm
Remote interfaces
GP-IB, RS-232, Ethernet (TCP/IP); socket and VXI-11 Ethernet modes
Bench size and mass
Approx. 426 × 221 × 459 mm (W × H × D), excluding protectors/handle; approx. 19 kg

Sales and Service

Warranty
6-Month Warranty
Calibration Status
Calibration certificate or test report available (as stated in your quotation)
Manufacturer-Listed Standard Accessories
Original manufacturer list: one regional power cord (none with -Y); two A9088ZM rubber feet supplied as one sheet; one A1023US manual CD; one each of the Getting Started Guide, China document, safety manual and worldwide contact list. Manufacturer-listed standard accessories are a reference to the original model documentation. They do not guarantee the contents of a used instrument shipment. Confirm the actual supplied items for the specific unit before ordering.
Downloads

Documents

Why work with us?

We check the details with you

From specifications to accessories, we help you understand what's included.

Know what you're buying

We share what we know about the condition, along with any available records.

Tell us what you need

Have a specific requirement or deadline? We'll focus on what matters to you.

No need to start over

We keep track of your questions and previous discussions, so it's easy to pick up where we left off.

A closer look at AQ6374

A closer look at AQ6374

Open the product inquiry form